top of page

These resistance sensors are designed for temperature measurement of gaseous substances in water-protected areas – e.g. for temperature measurement in rooms (schools, theatres, lecture halls, etc.), offices, interiors of residential houses or even production floors. Suitable design and high-quality material ensure that the sensors do not feel disturbing even in the interiors with high aesthetic requirements. You can choose the color of the box white or gray.

The temperature range of the sensors is -30 to 100 °C. The sensors meet ingress protection IP 30 according to EN 60529, as amended. Installation is recommended on an inner wall at the height of 1.5 m, in areas of movement of persons, at places not exposed to direct sunlight and not influenced by heat from walls, heating radiators or lighting.

The sensors are designed to be operated in a chemically non-aggressive environment, the use must be chosen with regard to the temperature and chemical resistance of the sensor head.

 

DECLARATION, CERTIFICATION, CALIBRATION

  • Manufacturer provides EU Declaration of Conformity.
  • Calibration – The final metrological inspection – comparison with standards or working instruments – is carried out for all the products. Continuity of the standards and working measuring instruments is ensured within the meaning of the Section 5 of Act no.505/1990 on metrology. The manufacturer offers a possibility to supply the sensors calibrated in SENSIT s.r.o.’s laboratory (according to requirements of the EN ISO/IEC 17025 standard, as amended) or in an Accredited laboratory.  
  • SPECIFICATIONS
  • DIMENSIONAL DRAFT
  • WIRING DIAGRAM
  • MODIFICATIONS
  • DOWNLOAD
Sensor typeNS 100NS 101NS 102NS 300NS 301
Type of sensing elementNi 1000/5000Ni 1000/6180Ni 891Ni 10000/5000Ni 10000/6180
Measuring range-30 to 100 °C LEXAN
Maximum measuring DC current1 mA1mA1mA0.3 mA0.3 mA

 

Sensor typeNS 103PTS 100PTS 200PTS 300HS 100
Type of sensing elementT1 = Ni 2226Pt 100/3850Pt 500/3850Pt 1000/3850thermistor NTC 20 kΩ
Measuring range-30 to 100 °C LEXAN
Maximum measuring DC current0.7 mA2 mA1.2 mA0.8 mA1 mW *)

*) maximum power consumption

Sensor typeNS 500NS 700Note
Type of sensing elementPt 1000/3850Pt 1000/3850 
Output signal4 to 20 mA0 to 10 V 
Measuring ranges-30 to 60 °C
0 to 35 °C
0 to 100 °C
0 to 150 °C
-30 to 60 °C
0 to 35 °C
0 to 100 °C
0 to 150 °C
Ambient temperature around the connection head:
-30 to 70 °C
Power supply (Ucc)12 to 30 VDC15 to 30 VDCRecommended value:
NS 500: 12 V DC
NS 700: 15 V DC
Max. load resistance Rs150 Ω for power supply 12 VDC
700 Ω for power supply 24 VDC
> 10 kΩ 
Sensing element break> 24 mA> 10.5 V 
Sensing element short< 3.5 mA~ 0 V 

**)According to the customer‘s requirement, it is possible to provide a customized measuring range from -40 to 150 °C; the minimum span of the range must be 35 °C (e.g. -20 to 15 °C; -30 to 80 °C)

OTHERS PARAMETRS
Accuracy classNi sensing elements: B class, Δt = ± (0.4 + 0.007t), for t ≥ 0; Δt = ± (0.4 + 0.028|t|), for t ≤ 0 in °C;
Pt sensing elements: B class according EN 60751, Δt = ± (0.3 + 0.005|t|) in °C
NTC 20 °C kΩ: ± 1 °C for the range 0 °C to 70 °C
Measuring error for NS 500 (range -30 °C to 70 °C)power supply = 12 V: ±(0.5 °C + 0.2 % from range)
power supply = 24 V: (-0.2 to 0.80 °C) ± 0.2 % from range
Measuring error for NS 700 (range -30°C to 70°C)power supply = 15 V: ±(0.5 °C + 0.2 % from range)
power supply = 24 V: (-0.2 to 0.80 °C) ± 0.2 % from range
Sensor connectionaccording to the wiring diagram
Time responseτ0.5 < 8 s (in air flow 0.4 m.s-1)
Recommended wire cross section0.35 to 1.5 mm2
Ingress protectionIP 30 in accordance with EN 60529
Material of the connection headLEXAN
Dimensions of connection head71.9 x 59 x 27 mm
Operating conditionsambient temperature:
-30 to 100 °C without converter
-30 to 70 °C with a converter
relative humidity: max 85 % (at the ambient temperature 25 °C)
atmospheric pressure: 87 to 107 kPa
Weight0.05 kg

IP30 חיישן טמפרטורה NS 500

0.00 ₪מחיר
    bottom of page