These resistance sensors are designed for temperature measurement of gaseous substances in water-protected areas – e.g. for temperature measurement in rooms (schools, theatres, lecture halls, etc.), offices, interiors of residential houses or even production floors. Suitable design and high-quality material ensure that the sensors do not feel disturbing even in the interiors with high aesthetic requirements. You can choose the color of the box white or gray.
The temperature range of the sensors is -30 to 100 °C. The sensors meet ingress protection IP 30 according to EN 60529, as amended. Installation is recommended on an inner wall at the height of 1.5 m, in areas of movement of persons, at places not exposed to direct sunlight and not influenced by heat from walls, heating radiators or lighting.
The sensors are designed to be operated in a chemically non-aggressive environment, the use must be chosen with regard to the temperature and chemical resistance of the sensor head.
DECLARATION, CERTIFICATION, CALIBRATION
- Manufacturer provides EU Declaration of Conformity.
- Calibration – The final metrological inspection – comparison with standards or working instruments – is carried out for all the products. Continuity of the standards and working measuring instruments is ensured within the meaning of the Section 5 of Act no.505/1990 on metrology. The manufacturer offers a possibility to supply the sensors calibrated in SENSIT s.r.o.’s laboratory (according to requirements of the EN ISO/IEC 17025 standard, as amended) or in an Accredited laboratory.
- SPECIFICATIONS
- DIMENSIONAL DRAFT
- WIRING DIAGRAM
- MODIFICATIONS
- DOWNLOAD
Sensor type | NS 100 | NS 101 | NS 102 | NS 300 | NS 301 |
Type of sensing element | Ni 1000/5000 | Ni 1000/6180 | Ni 891 | Ni 10000/5000 | Ni 10000/6180 |
Measuring range | -30 to 100 °C LEXAN | ||||
Maximum measuring DC current | 1 mA | 1mA | 1mA | 0.3 mA | 0.3 mA |
Sensor type | NS 103 | PTS 100 | PTS 200 | PTS 300 | HS 100 |
Type of sensing element | T1 = Ni 2226 | Pt 100/3850 | Pt 500/3850 | Pt 1000/3850 | thermistor NTC 20 kΩ |
Measuring range | -30 to 100 °C LEXAN | ||||
Maximum measuring DC current | 0.7 mA | 2 mA | 1.2 mA | 0.8 mA | 1 mW *) |
*) maximum power consumption
Sensor type | NS 500 | NS 700 | Note |
Type of sensing element | Pt 1000/3850 | Pt 1000/3850 | |
Output signal | 4 to 20 mA | 0 to 10 V | |
Measuring ranges | -30 to 60 °C 0 to 35 °C 0 to 100 °C 0 to 150 °C | -30 to 60 °C 0 to 35 °C 0 to 100 °C 0 to 150 °C | Ambient temperature around the connection head: -30 to 70 °C |
Power supply (Ucc) | 12 to 30 VDC | 15 to 30 VDC | Recommended value: NS 500: 12 V DC NS 700: 15 V DC |
Max. load resistance Rs | 150 Ω for power supply 12 VDC 700 Ω for power supply 24 VDC | > 10 kΩ | |
Sensing element break | > 24 mA | > 10.5 V | |
Sensing element short | < 3.5 mA | ~ 0 V |
**)According to the customer‘s requirement, it is possible to provide a customized measuring range from -40 to 150 °C; the minimum span of the range must be 35 °C (e.g. -20 to 15 °C; -30 to 80 °C)
OTHERS PARAMETRS | |
Accuracy class | Ni sensing elements: B class, Δt = ± (0.4 + 0.007t), for t ≥ 0; Δt = ± (0.4 + 0.028|t|), for t ≤ 0 in °C; Pt sensing elements: B class according EN 60751, Δt = ± (0.3 + 0.005|t|) in °C NTC 20 °C kΩ: ± 1 °C for the range 0 °C to 70 °C |
Measuring error for NS 500 (range -30 °C to 70 °C) | power supply = 12 V: ±(0.5 °C + 0.2 % from range) power supply = 24 V: (-0.2 to 0.80 °C) ± 0.2 % from range |
Measuring error for NS 700 (range -30°C to 70°C) | power supply = 15 V: ±(0.5 °C + 0.2 % from range) power supply = 24 V: (-0.2 to 0.80 °C) ± 0.2 % from range |
Sensor connection | according to the wiring diagram |
Time response | τ0.5 < 8 s (in air flow 0.4 m.s-1) |
Recommended wire cross section | 0.35 to 1.5 mm2 |
Ingress protection | IP 30 in accordance with EN 60529 |
Material of the connection head | LEXAN |
Dimensions of connection head | 71.9 x 59 x 27 mm |
Operating conditions | ambient temperature: -30 to 100 °C without converter -30 to 70 °C with a converter relative humidity: max 85 % (at the ambient temperature 25 °C) atmospheric pressure: 87 to 107 kPa |
Weight | 0.05 kg |